Intraocular Pressure Screening
Intraocular pressure (IOP) measurement is one of the most performed procedures by the CPOA. IOP is a critical parameter in glaucoma screening, diagnosis, and management. While IOP measurement alone cannot diagnose or exclude glaucoma (the optic nerve and visual fields are equally important), accurate IOP recording is foundational to monitoring glaucoma patients and identifying those at risk. The CPOA must be skilled in at least one tonometry method and understand the factors that affect IOP accuracy.
Normal IOP
Normal IOP is generally considered to be 10-21 mmHg, with a statistical mean of approximately 15.5 mmHg in the general population. However:
- Some patients develop glaucoma with normal IOP (normal-tension glaucoma)
- Many patients have IOP above 21 mmHg without developing glaucoma (ocular hypertension)
- IOP fluctuates diurnally -- typically highest in the morning (upon waking) and lowest at night
- The target IOP for a treated glaucoma patient is set individually by the physician
Non-Contact Tonometry (NCT / Air Puff)
Non-contact tonometry uses a brief, calibrated puff of air to flatten (applanate) the central cornea. The instrument measures the time or force required for applanation. No drops are required, making it suitable for screening and high-volume settings. The CPOA commonly performs NCT as a preliminary test.
NCT Technique
- Seat the patient at the instrument; adjust the chin rest
- Instruct the patient to keep both eyes open and look at the target light inside the instrument
- Align the instrument with the patient's corneal reflex
- Capture 3 readings per eye and record the average
- Do not touch the eyelids to hold them open -- this alters IOP by applying pressure
💡 Clinical Tip: NCT readings can be unreliable if the patient blinks or squeezes during the air puff, or if the alignment is off-center. Take 3 readings per eye; if there is significant variability (more than 3-4 mmHg between readings), note this and consider Goldmann applanation tonometry for more accurate assessment.
Goldmann Applanation Tonometry (GAT)
GAT is the gold standard for IOP measurement. It is performed at the slit lamp using the prism and fluorescein to directly measure the force required to flatten a defined area (3.06 mm diameter) of the corneal surface.
GAT Technique
- Instill topical anesthetic (proparacaine) to minimize discomfort
- Instill fluorescein (from a fluorescein strip or sodium fluorescein drops)
- Switch to the cobalt blue filter on the slit lamp
- Advance the tonometer prism until it makes contact with the cornea
- Observe the two fluorescein semicircles through the eyepiece
- Adjust the force dial until the inner edges of the two semicircles just touch ("kiss" appearance)
- Read the value on the drum and multiply by 10 to get mmHg (if the drum scale reads in gram-force units) -- or read directly in mmHg depending on the scale
- Record the reading; repeat for the fellow eye
The Fluorescein Mires
When properly aligned, two semicircles (mires) are visible through the prism -- one above and one below. The clinician adjusts the force drum until:
- The mires are equal in height (if unequal, the prism may be off-axis or corneal astigmatism affects the appearance)
- The inner edges of the two mires just touch (not overlapping, not separated)
⚠️ Common Mistake: If the mires are too thin (not enough fluorescein) they may be difficult to see; if too thick (too much fluorescein), the IOP reading will be falsely low. The mires should appear as two bright green semicircles, each with a width roughly one-tenth of their radius.
Factors Affecting IOP Accuracy
| Factor | Effect on IOP Reading |
|---|---|
| Thick cornea (CCT >570 microns) | Falsely HIGH reading |
| Thin cornea (CCT <520 microns) | Falsely LOW reading |
| Blepharospasm / lid squeezing | Falsely HIGH (external pressure on globe) |
| Breath-holding by patient | Falsely HIGH (Valsalva increases IOP) |
| High corneal astigmatism (>3 D) | Inaccurate unless prism rotated to 43 degrees |
| Fluorescein too thick / too thin | Falsely LOW or difficult to read |
| Patient staring; tight collar / tie | Falsely HIGH |
| Post-LASIK cornea (thinned) | Falsely LOW |
Tono-Pen and Rebound Tonometer
Tono-Pen: a handheld electronic applanation device; useful for patients who cannot be positioned at the slit lamp (wheelchair-bound, children, hospital patients). Readings are less precise than GAT.
Rebound tonometer (iCare): uses a small probe that bounces off the cornea; requires no drops or contact lens; particularly useful in children or uncooperative patients. Also handheld.
IOP Measurement in Glaucoma Management
In glaucoma patients, IOP is measured at every visit to monitor response to treatment. The CPOA records the reading, the time of measurement (diurnal variation is significant), and which medications the patient took before the visit. A significant rise from baseline (e.g., >3-4 mmHg) or an IOP above the physician-set target triggers physician review.
Key Takeaways
- Normal IOP is 10-21 mmHg; mean is ~15.5 mmHg; diurnal variation peaks in the morning
- NCT (air puff) is the standard screening method; take 3 readings and average
- Goldmann applanation tonometry is the gold standard; uses fluorescein + cobalt blue; mires' inner edges should just touch
- Thick cornea = falsely high IOP; thin cornea = falsely low IOP; CCT must be considered in interpretation
- Lid squeezing, Valsalva, and tight clothing around the neck falsely elevate IOP
- Document IOP readings with the time taken; diurnal variation is clinically significant in glaucoma