Loading...
Loading...
Goldmann applanation tonometry (GAT) is the gold standard method for measuring intraocular pressure (IOP). It is the reference technique against which all other tonometers are calibrated, and it appears consistently on the COA exam within the Assessments domain. Understanding its physical principle, setup procedure, mire alignment endpoint, and sources of error is essential for both the exam and clinical practice.
Unlike non-contact tonometry, GAT requires topical anesthesia and fluorescein instillation, direct contact with the cornea, and careful technique to obtain accurate readings. For COA candidates, mastering GAT means understanding not just what to do, but why each step matters and how errors affect results.
GAT operates on the Imbert-Fick principle: the pressure inside a thin-walled sphere equals the force required to flatten (applanate) its surface divided by the area flattened. Goldmann refined this for the eye by discovering that when exactly a 3.14 mm diameter circle of cornea is flattened, two opposing forces — tear film surface tension (which pulls the prism toward the eye) and corneal rigidity (which pushes back) — precisely cancel each other out. This elegant cancellation makes the measurement largely independent of these confounders for a cornea of average thickness and curvature.
The force (in grams) needed to achieve this exact flattening area is read from the calibration drum and multiplied by 10 to convert to mmHg. So a drum reading of 1.6 corresponds to an IOP of 16 mmHg.
3.14 mm
Diameter flattened at endpoint
10–21 mmHg
Statistical norm (95th percentile)
520–540 µm
Calibrated for average CCT
Goldmann determined through experimentation that at exactly 3.14 mm of applanation diameter, the error introduced by surface tension (which causes overestimation) is exactly offset by the error introduced by corneal rigidity (which causes underestimation). This sweet spot makes the measurement self-correcting for a cornea of average biomechanical properties.
Proper setup before the measurement is as important as the measurement itself. Errors in preparation — wrong positioning, too much fluorescein, skipping calibration — degrade accuracy before the prism ever touches the cornea.
Before each clinical session, check the tonometer calibration with the calibration bar. At the 0 setting, the arm should be in neutral; with the calibration weight applied, the arm should deflect to the appropriate mark. A tonometer out of calibration by more than 0.5 mmHg must not be used clinically until serviced.
Wipe the prism with an alcohol-based disinfectant and allow it to air dry completely (at least 5 minutes) before patient contact. Never use a wet prism — residual disinfectant on the eye causes corneal epithelial toxicity. Disposable prisms eliminate this concern entirely and are preferred when disease transmission is a risk.
Apply one drop of topical anesthetic (proparacaine 0.5% or tetracaine 0.5%) to each eye and wait 30 seconds for full effect. Then apply fluorescein: either a fluorescein strip moistened with one drop of anesthetic, or a combination fluorescein-anesthetic drop (e.g., Fluress). The goal is a thin, even fluorescein layer — enough to create visible mires, but not so much that mires appear excessively thick.
Seat the patient comfortably at the slit lamp with the chin in the chin rest and forehead against the forehead strap. Ask the patient to fixate on a distant target with the eye not being tested to minimize accommodation-driven eye movement. Set the cobalt blue filter on the slit lamp illumination arm. Ask the patient to breathe normally — Valsalva (breath-holding) significantly raises IOP.
Set the drum to approximately 1.0 (10 mmHg) as a starting point. Turn on the cobalt blue light. Gently advance the joystick to bring the prism into contact with the central cornea while watching from the side. Once contact is made, view through the slit lamp oculars in low magnification to see the two fluorescent semicircles.
The split prism of the GAT head creates two fluorescent semicircles (mires) when the prism contacts fluorescein-stained tear film. The measurement endpoint is reached when the inner borders of these two semicircles just touch — not overlapping, not separated. This visual endpoint is called inner-border touching or inner-edge alignment.
Inner borders have a gap between them. Force is too low.
Action: Increase dial (more force)
The two semicircles just make contact at their inner edges. This is the endpoint.
Action: Read the drum value
Inner borders cross each other. Force is excessive.
Action: Decrease dial (less force)
In eyes with corneal astigmatism greater than 3 diopters, the mires become elliptical rather than circular, introducing measurement error. The correction: rotate the prism so that the red axis mark aligns with the minus cylinder axis of the patient's refraction. This orientation averages the IOP across the two principal meridians and reduces the astigmatic error significantly.
Because GAT is calibrated for an average CCT of approximately 520–540 µm, eyes with non-average CCT introduce systematic error. This is one of the most tested concepts related to GAT on the COA exam and is clinically critical for glaucoma risk stratification.
| CCT Value | GAT Effect | Approximate Error | Clinical Implication |
|---|---|---|---|
| Thin (<500 µm) | Underestimates true IOP | ~1 mmHg per 20 µm below average | May miss elevated IOP; glaucoma risk underestimated |
| Average (520–540 µm) | Accurate reading | No correction needed | GAT reading is reliable |
| Thick (>560 µm) | Overestimates true IOP | ~1 mmHg per 20 µm above average | May overdiagnose ocular hypertension |
| Post-LASIK cornea | Significantly underestimates | 2–5 mmHg depending on ablation | Requires correction formula; GAT alone unreliable |
Valsalva Maneuver
Breath-holding raises episcleral venous pressure, which elevates measured IOP. Instruct patients to breathe normally throughout the measurement. A difference between readings may indicate the patient is holding their breath.
Lid Squeezing (Blepharospasm)
When patients squeeze their eyelids, orbicularis muscle contraction raises IOP. Use a lid speculum if necessary or have the patient open their eye as wide as possible. Never force the lids open with your fingers as this directly transmits pressure to the globe.
Excessive or Insufficient Fluorescein
Too much fluorescein causes thick mires (overestimation of applanation area, underestimation of IOP). Too little fluorescein makes mires hard to see, leading to inaccurate endpoint identification. Aim for mire width of about one-tenth the diameter of the mire ring.
Corneal Edema
Edematous corneas are softer than normal and require less force to applanate, leading to IOP underestimation. This is particularly relevant in patients with Fuchs dystrophy or after intraocular surgery. The noncontact tonometer or Tono-Pen may also be inaccurate in edematous corneas.
Tight Neckwear / Collar
A tight collar or tie compresses the jugular veins, impeding episcleral venous drainage and transiently raising IOP by 2–4 mmHg. Ask patients to loosen tight collars before measurement.
Accurate documentation of IOP measurements is both a clinical and medicolegal requirement. The COA exam tests documentation format, and errors in recording can create significant clinical problems in ongoing glaucoma management.
Include OD and OS values
Always document both eyes, even if only one has pathology. Example: IOP OD 16 mmHg, OS 18 mmHg by GAT.
Record time of measurement
IOP fluctuates throughout the day (diurnal variation), typically peaking in the early morning. Time of measurement is essential for comparing serial readings accurately.
Specify the tonometer type
Document whether the reading was obtained by GAT, NCT, iCare, or another method. Different instruments have different accuracy profiles and are not directly interchangeable.
Note medications affecting IOP
Topical beta-blockers, prostaglandin analogs, and other glaucoma medications lower IOP. Steroids raise it. Document current ophthalmic medications when recording IOP.
Opterio includes Goldmann tonometry questions within COA Assessments domain practice, with AI-powered explanations that reinforce the clinical reasoning behind each answer.
Statistical norms, diurnal variation, ocular hypertension, and clinical significance.
Air puff technique, advantages, limitations, and when to confirm with GAT.
GAT, NCT, iCare, Tonopen, and Perkins compared for accuracy and indications.
Exam format, content domains, eligibility, pass rates, and registration.
Goldmann applanation tonometry is based on the Imbert-Fick law, which states that for an ideal thin-walled dry sphere, the pressure inside equals the force needed to flatten the surface divided by the area flattened. GAT applies just enough force to flatten a standardized 3.14 mm diameter area of the cornea. At that exact area, tear film surface tension and corneal rigidity cancel each other out, making the measurement accurate for a normal cornea. The force required to achieve this flattening, read on the calibration drum, is multiplied by 10 to yield IOP in mmHg.
The tonometer prism creates two fluorescent semicircles when viewed through the cobalt blue light. You rotate the dial to increase or decrease the force until the inner edges of the two semicircles just touch each other — not overlapping and not with a gap. This endpoint is called inner-border touching. If the mires are too thin (inner borders separated), increase the force. If the mires are too thick (inner borders overlapping), decrease the force. The mires should also be of equal width; if one appears thicker, the prism may be tilted or decentered.
Standard GAT assumes a central corneal thickness (CCT) of approximately 520–540 µm. For every 10 µm thicker than average, IOP is overestimated by approximately 0.5 mmHg; for every 10 µm thinner, IOP is underestimated by the same amount. A patient with a thin cornea (e.g., 480 µm) may have a true IOP higher than the tonometer reads, which is clinically important for glaucoma risk assessment. Pachymetry is performed to measure CCT and allow the clinician to interpret the GAT reading in context.
You need both a topical anesthetic and fluorescein for GAT. The anesthetic (proparacaine or tetracaine) eliminates patient discomfort and blink reflex, allowing the prism to contact the cornea steadily. Fluorescein stains the tear film and fluoresces under cobalt blue illumination, creating the two semicircular mires needed to identify the applanation endpoint. Combination anesthetic-fluorescein strips (Fluress) or separate instillation of drops followed by a fluorescein strip are both acceptable. Too much fluorescein makes the mires too thick; too little makes them difficult to see.
Key sources of error include: (1) Valsalva maneuver — breath-holding raises episcleral venous pressure and temporarily elevates IOP; instruct patients to breathe normally. (2) Tight collars or neck pressure on jugular veins can elevate IOP. (3) Corneal astigmatism greater than 3 diopters causes mires that are not perfectly circular; the prism should be rotated so the negative axis bisects the two mires. (4) Corneal edema softens the cornea, causing underestimation. (5) Excessive fluorescein creates thick mires and overestimation. (6) Patient squeezing (lid squeezing elevates IOP from orbicularis pressure). (7) Calibration error if the instrument has not been checked recently.
Calibration is checked using the tonometer's built-in calibration bar or a separate calibration weight set. Set the dial to 0 mmHg and attach the test weight that corresponds to a known pressure (typically 0, 2, and 6 diopter equivalents). The tonometer arm should move appropriately for each weight, indicating the spring mechanism is within tolerance. If the tonometer is out of calibration by more than 0.5 mmHg at any check point, it must be sent for service before clinical use. Calibration should be performed at regular intervals per manufacturer recommendations and whenever results seem clinically inconsistent.